Determination of Carbamazepine in Human Urine and Serum Samples by High‐performance Liquid Chromatography with Post‐column Ru(bipy)‐Ce(SO4)2 Chemiluminescence Detection

Xing‐yu Long,Fu‐nan Chen
DOI: https://doi.org/10.1002/bio.2365
IF: 2.613
2012-01-01
Luminescence
Abstract:ABSTRACTA novel, sensitive and rapid CL method coupled with high‐performance liquid chromatography separation for the determination of carbamazepine is described. The method was based on the fact that carbamazepine could significantly enhance the chemiluminescence of the reaction of cerium sulfate and tris(2,2‐bipyridyl) ruthenium(II) in the presence of acid. The chromatographic separation was performed on a Kromasil® (Sigma‐Aldrich) TM RP‐C18 column (id: 150 mm × 4.6 mm, particle size: 5 µm, pore size: 100 Å) with a mobile phase consisting of methanol–water‐glacial acetic acid (70:29:1, v/v/v) at a flowrate of 1.0 mL/min, the total analysis time was within 650 s. Under optimal conditions, CL intensity was linear for carbamazepine in the range 2.0 × 10−8 ~ 4.0 × 10−5 g/mL, with a detection limit of 6.0 × 10−9 g/mL (S/N = 3) and the relative standard detection was 2.5% for 2.0 × 10−6 g/mL (n = 11). This method was successfully applied to the analysis of carbamazepine in human urine and serum samples. The possible mechanism of the CL reaction is also discussed briefly. Copyright © 2012 John Wiley & Sons, Ltd.