Capillary Electrophoresis with Electrochemiluminescence Detection of Procyclidine in Human Urine Pretreated by Ion-Exchange Cartridge

XH Sun,JF Liu,WD Cao,XR Yang,EK Wang,YS Fung
DOI: https://doi.org/10.1016/s0003-2670(02)00780-8
IF: 6.911
2002-01-01
Analytica Chimica Acta
Abstract:This paper describe a Ru(bpy)32+ based electrochemiluminescence (ECL) method to detect procyclidine in human urine following separation by capillary electrophoresis (CE). An ECL detection cell was designed for post-column addition of Ru(bpy)32+. Parameters affecting separation and detection were optimized, leading to a detection limit of 1×10−9mol/l in an on-capillary stacking mode. For application in urine, a cartridge packed with slightly acidic cation-exchange resin was used to eliminate the matrix effects of urine and improve the detection sensitivity. Extraction recovery was nearly 90%.
What problem does this paper attempt to address?