Ion chromatography combined with online electrochemical derivatization and fluorescence detection for the determination of carbamazepine in human plasma.

Shuchao Wu,Wei Xu,Qamar Subhani,Bingcheng Yang,Deying Chen,Yan Zhu,Lanjuan Li
DOI: https://doi.org/10.1016/j.talanta.2012.09.039
IF: 6.1
2012-01-01
Talanta
Abstract:This paper describes the determination of carbamazepine (CBZ) in human plasma using ion chromatography combined with online electrochemical derivatization and fluorescence detection. Separation of CBZ with anion exchange column was demonstrated to be feasible using either basic (10mM NaOH) or acidic (0.1 M H3PO4) reagent with a small amount of acetonitrile (ACN) added as eluent. Electrochemical derivatization of CBZ into a strongly fluorescent product, which could be carried out only under the acidic condition, was investigated via the previously reported electrolytic cell (EC), as well as two modes of acidification. The linear range of CBZ for human plasma was between 10–2000μgL−1 under the optimized experimental conditions. The limit of detection (LOD, S/N=3) was 1.3μgL−1 and the relative standard deviation (RSD, n=7) was 2.6%. Better sensitivity and selectivity of the present method were demonstrated in comparison with ion chromatography with ultraviolet detection (IC-UV). The spiked recoveries of CBZ in 2 human plasma samples were 78.5 and 114%, respectively.
What problem does this paper attempt to address?