Improving the Sensitivity of DC Magneto-Optical Kerr Effect Measurement to 10-7 Rad/hz

Junying Ma,Feng Gu,Ying Xu,Jiaming Le,Fanlong Zeng,Yizheng Wu,Chuanshan Tian
DOI: https://doi.org/10.3788/col202220.111201
IF: 2.56
2022-01-01
Chinese Optics Letters
Abstract:A high-sensitivity DC magneto-optical Kerr effect (MOKE) apparatus is described in this Letter. Via detailed analysis on several dominating noise sources, we have proposed solutions that significantly lower the MOKE noise, and a sensitivity of 1.5×10-7 rad/Hz is achieved with long-term stability. The sensitivity of the apparatus is tested by measuring a wedge-shaped Ni thin film on SiO2 with Ni thickness varying from 0 to 3 nm. A noise floor of 1.5×10-8 rad is demonstrated. The possibility of further improving sensitivity to 10-9 rad via applying AC modulation is also discussed.
What problem does this paper attempt to address?