Conditions for high efficient sorting experiments by flow cytometry

徐琴,黄莹莹,姜宇
DOI: https://doi.org/10.3760/cma.j.issn.1007-1245.2021.01.006
2021-01-01
Abstract:Objective:To improve the yield and purity of sorting experiments by flow cytometry, to establish efficient sorting conditions, and to provide references for the application of fluorescence-activated cell sorting.Methods:The beads were used as the test samples, and the BD FACS Sorp Aria II was used as the test instrument from October, 2019 to June, 2020. Different sorting velocities, sorting modes, and nozzles were investigated, and the influencing factors were analyzed. The na?ve T cells were sorted under the optimized conditions.Results:The BD FACS Sorp Aria II gained purity of more than 98% with four different sorting velocities, flow rates of 1.0, 3.0, 5.0, and 7.0, by using a 70 μm nozzle and the purity mode. However, the higher the flow rate, the lower the sorting yield. Sorting samples by different sorting modes, the efficiency and yield of the yield sorting mode were (99.67±0.33)% and (79.78±7.14)%, respectively, which were higher than those of the other two modes, but the purity was the lowest one by the yield sorting mode (both P<0.05). The instrument gained a purity of more than 98% with all the three nozzles, and the yields of the three nozzles were not statistically different ( P>0.05). Conclusion:When na?ve T cells are sorted by BD FACS Sorp Aria II flow cytometer with a 70 μm nozzle, a flow rate of 1.0, and the purity sorting mode are preferred.