Performance Testing And Analysis Of Electrophoresis Process Driven By Low Voltage On The Soi Microchip

Liang Jing,Xu Yi,Hu Xiaoguo,Lu Jiali
2009-01-01
Abstract:In this paper, a new approach for amino acids separation on the electrophoresis chip based low voltage-driven electrophoresis was reported in detail. The insulation of the neighboring arrayed-electrodes was indicated by measuring breakdown voltage, which tested as 260V; by means of current monitoring method, the electroosmotic flow (EOF) of SOI-PDMS hybrid electrophoresis microchip was 3.9x10(-4)cm(2).V-1.s(-1); the range of working voltages which selected by measuring the Volt-Ampere characteristic curve was 100 similar to 140V.cm(-1). Using the laser-induced fluorescence detector, a set of low-voltage driven electrophoresis separating and detecting system was finally established. Under the optimized conditions, the mixture of FITC-labeled arginine(Arg) and phenylalanine (Phl)was successfully separated with a resolution of 3.14 on the low-voltage driven electrophoresis microchip.
What problem does this paper attempt to address?