Sensitive Detection Of H2o2 By Electrochemiluminescence (Ecl) Of Luminol With A Microfluidic Chip Flow Electrolysis Cell

Fang Fang,Zhi-Yong Wu,Yan-Yan Wen,Hubert H. Girault
2007-01-01
Abstract:Sensitive detection of H2O2 by Luminol electrochemiluminescence (ECL) is presented taking advantage of a new microfluidic flow electrolysis cell of only 300nl volume. Both cathode and anode were positioned parallel along the 2 sides of the channel. One ECL peak around 0.7 V vs. Pt pseudo reference was observed in a CV scan with 2-electrode configuration. It was confirmed that electrochemical oxidation of both luminol and H2O2 to radicals occurred on the anode, and photons were produced during luminol chemiluminescence (CL) reaction in homogeneous phase near anode; on the other hand, photon production through 02 route subjected to extra reaction steps, involving consumption of luminol radical liquid phase near anode to form a oxygen radical, and the oxygen radical reacted with another luminol radical to form the intermediate product to release photons, which explains the low efficiency of luminol ECL with 02 comparing to that of H2O2. Detection limit of H2O2 down to pM level was obtained taking advantage of the enhancement effect on luminol ECL with the flow cell.
What problem does this paper attempt to address?