Comparison Of Laser Phacovaporization Using The Er-Yag And The Er-Ysgg Laser

Rp Gailitis,Sw Patterson,Ma Samuels,K Hagen,Qs Ren,Go Waring
DOI: https://doi.org/10.1001/archopht.1993.01090050131042
1993-01-01
Archives of Ophthalmology
Abstract:Objective.-To study the interaction of the erbium (Er)-YAG (2.94 mum) and the Er-YSGG (2.79 mum) lasers with the human crystalline lens tissue.Methods.-Fresh human crystalline lens sections were used to measure the photovaporization threshold, rate, and damage zone of the two lasers.Results.-We found the photovaporization threshold for the Er-YAG and the ErYSGG lasers to be 1.4 J/cm2 and 5.5 J/cm2, respectively. At 10 J/cm2, the photovaporization rate is 67.9 mum per pulse for the Er-YAG laser and 18.4 mum per pulse for the Er-YSGG laser. The increased rate of photovaporization as a function of radiant energy for the Er-YAG laser is almost twice that for the Er-YSGG. Damage zones for the Er-YAG laser ranged from 4 to 9 mum compared with 10 to 22 mum for the Er-YSGG.Conclusions.-It is apparent that both lasers can adequately photovaporize human crystalline lens tissue and should be further studied for this purpose.
What problem does this paper attempt to address?