Determination Of Iron, Nickel, Copper, Sodium In Crude Oil And Residue Oil By Microwave Digestion-Microwave Plasma Torch Atomic Emission Spectrometry

Js Zhang,Lh Li,Qh Jin
IF: 1.193
2005-01-01
Chinese Journal of Analytical Chemistry
Abstract:The microwave digestion procedures were developed for crude oil and residue oil. Microwave plasma torch atomic emission spectrometry (MPT-AES) was used to determine the iron, nickel, copper, sodium in crude oil and residue oil. Optimal condition for determination ( analytical wavelength, microwave power, flow rate of carrier gas, flow rate of supporting gas and acid concentrations) were chosen. The effect of concomitant elements on determination of iron, nickel, copper, sodium were studied. The detection limits were 22, 42, 2.0 and 1.0 mu g/L for iron, nickel, copper, sodium, respectively. The linear range were 0.1 similar to 100, 0.15 similar to 50, 0.01 similar to 5.0 and 0.006 similar to 2.0 mg/L, respectively. The relative standard deviation for the determination of samples were less than 3.9%, 3.6%, 5.5.% , 3.4% respectively. The determination results of microwave digestion were identical with those obtained with routine dry-ash method. The results show that the method is simple, rapid, accurate, save reagent, less environment contamination and of great application value.
What problem does this paper attempt to address?