Time-Resolved In Situ Detection Of Co In A Shock Tube Using Cavity-Enhanced Absorption Spectroscopy With A Quantum-Cascade Laser Near 4.6 Mu M

Kai Sun,Shengkai Wang,Ritobrata Sur,Xing Chao,Jay B. Jeffries,Ronald K. Hanson
DOI: https://doi.org/10.1364/OE.22.024559
IF: 3.8
2014-01-01
Optics Express
Abstract:Cavity-enhanced absorption spectroscopy (CEAS) using a mid-infrared DFB quantum-cascade laser is reported for sensitive time-resolved (10 mu s) in situ CO measurements in a shock tube. Off-axis alignment and fast scanning of the laser wavelength were used to minimize coupling noise in a low-finesse cavity. An absorption gain factor of 91 was demonstrated, which enabled sub-ppm detection sensitivity for gas temperatures of 1000-2100K in a 15 cm diameter shock tube. This substantial improvement in detection sensitivity compared to conventional single-pass absorption measurements, shows great potential for the study of reaction pathways of high-temperature combustion kinetics mechanisms in shock tubes. (C)2014 Optical Society of America
What problem does this paper attempt to address?