Christopher A. Curwen,Mathieu Fradet,Ryan M. Briggs
Abstract:We present a modified approach to laser optical-feedback cavity-enhanced absorption spectroscopy. The technique involves continuously scanning the length of a high-finesse cavity to periodically lock a diode laser to the cavity resonance, resulting in a discrete set of transmission measurements that are evenly spaced in frequency. For a fixed laser bias, data can be collected spanning a spectral bandwidth equivalent to the free-spectral range of the cavity, with spectral resolution inversely proportional to the distance from the laser to cavity. The center frequency of this scan can be tuned by tuning the free-running laser frequency. We demonstrate the concept using a fiber-coupled 1578-nm laser and a scanning Fabry-Perot cavity to measure a series of weak CO2 absorption lines with a frequency resolution of 15.6 MHz and a noise equivalent absorption coefficient of 10-7 cm-1, limited by the moderate finesse (~5000) and short length (~5 cm) of the cavity. Individual CO2 line shapes can be measured with high resolution in a single scan that takes 67 ms. The approach has a combination of characteristics that are advantageous for in situ instruments, such as small size, high spectral resolution, fast data collection, and minimal components.
What problem does this paper attempt to address?
The problem this paper attempts to address is improving the performance of laser absorption spectroscopy technology in terms of rapid, high-resolution data acquisition, particularly in the miniaturization and simplification of complexity for field instrument applications. Specifically, the authors propose an improved Optical Feedback Cavity Enhanced Absorption Spectroscopy (OF-CEAS) technique. By continuously scanning the length of a high-finesse cavity, a semiconductor laser is periodically locked to the cavity's resonance frequency, generating a series of frequency-uniformly distributed transmission data points. This method enables the rapid collection of high-resolution spectral data in a single scan without the need for complex locking or feedback loops.
### Main Issues:
1. **Improving Spectral Resolution**: Traditional OF-CEAS technology is limited by the cavity's Free Spectral Range (FSR) and can only collect data at fixed spectral points. The method proposed in this paper collects high-resolution data over a larger spectral range by scanning the cavity length.
2. **Rapid Data Acquisition**: Traditional high-resolution spectroscopy techniques usually require a long time to complete measurements. The method proposed in this paper can quickly collect data in a single scan, with a scan time of only 67 milliseconds.
3. **Miniaturization and Simplification**: Traditional high-sensitivity spectroscopy techniques often require complex equipment and stable systems. The method proposed in this paper uses a short cavity and minimized components, making it suitable for field instrument applications.
### Solution:
- **Scanning Cavity Technique**: By using a piezoelectric actuator to scan the length of a high-finesse cavity, the laser is periodically locked to the cavity's resonance frequency, generating uniformly distributed transmission data points.
- **Fixed Laser Bias**: With a fixed laser bias, spectral data spanning the cavity's Free Spectral Range can be collected, with spectral resolution inversely proportional to the distance from the laser to the cavity.
- **Rapid Scanning**: The scanning speed is only limited by the laser locking time, allowing scanning at a frequency of 15 Hz, collecting about 1500 data points per second.
- **High-Resolution Measurement**: Using a 1578 nm fiber-coupled laser and a scanning Fabry-Perot cavity, a series of weak CO2 absorption lines were measured with a frequency resolution of 15.6 MHz and a noise-equivalent absorption coefficient of 10^-7 cm^-1.
### Experimental Results:
- **High-Resolution Spectra**: Experimental results show that high-resolution CO2 absorption line shapes can be measured in a single scan, with a scan time of 67 milliseconds.
- **Data Processing**: By averaging overlapping data points, the signal-to-noise ratio can be further improved, obtaining more accurate absorption spectra.
- **Sensitivity**: Under cavity conditions, the average intensity fluctuation is 10^-3, corresponding to a minimum detectable absorption coefficient of about 10^-7 cm^-1. Although this is still behind the sensitivity of state-of-the-art cavity-enhanced systems, the method has proof-of-concept significance.
### Conclusion:
This paper proposes an improved OF-CEAS technique that achieves rapid, high-resolution spectral data acquisition by scanning the length of a high-finesse cavity. The method features miniaturization, high resolution, rapid data acquisition, and simple components, making it suitable for field instrument applications. Future work can further improve sensitivity by using higher reflectivity mirrors and longer cavities.