Multi-angle Scheimpflug Projection 3D Microscope: Design, Calibration, and Three-Dimensional Reconstruction
Huiwen Deng,Yong Liu,Guofeng Zhang,Xing Qu,Lingyu Ge,Shuming Yang
DOI: https://doi.org/10.1016/j.measurement.2023.113609
IF: 5.6
2023-01-01
Measurement
Abstract:3D measurement of complex micro-structures has becoming increasingly important for precision manufacturing. In this paper, a multi-angle Scheimpflug projection 3D microscope (MASP-3DM) is proposed, and the detailed design procedure is given. The system deployed four Scheimpflug projections and one telecentric imaging to overcome the shortcomings of limited depth of field (DOF), occlusion and reflection. Then, an improved and practical system calibration method is proposed, which utilizes pinhole projection model to calibrate the telecentric imaging. Meanwhile, local homography is combined with perspective-n-point (PnP) strategy to improve the calibration accuracy. Further, a point-cloud registration error correction model is established to achieve complete and high-precision measurement. The experimental results indicate that the re-projection error is within +/- 0.5 mu m and the registration error is less than 2 mu m for system calibration, the maximum of mean deviation, measurement bias, standard deviation, and type A uncertainty are respectively less than 3 mu m, 7 mu m, 4 mu m, and 1 mu m for 3D measurement, within the 10(H)mmx50(W)mmx40(D)mm volume.