Improving Accuracy of Averaging Pitot Tube Flow Sensor at Downstream of Elbow

Lijun Sun,Chao Sun,Xiang Qiu
DOI: https://doi.org/10.11784/tdxbz201209048
2014-01-01
Abstract:When the averaging Pitot tube flow sensor is installed at the downstream of elbow,the flow regime is not fully-developed turbulence,which will increase the error of flow measurement. In order to improve the accuracy of flow meas-urement,a method to reduce the error by installing a flow conditioner at the downstream of the 90°elbow for reforming the symmetry of flow velocity profile was put forward. Based on the NEL flow conditioner,the improved flow conditioner was designed by the method of computational fluid dynamics(CFD)simulation and analysis combined with real flow experiment. Verified by the simulation and experiment,after installing the improved flow conditioner,the linearity error of the averag-ing Pitot tube flow sensor was decreased to 0.176%. It has been found through analyzing the flow mechanism that the sym-metry of the velocity profile at the upstream of the flow sensor is the main factor affecting the error of the sensor. It is indi-cated in the data that the difference of the linearity error predicted by the CFD simulation is less than 0.056%.