Determination of Selenium in Marine Organism Samples by HG-AFS

Zhaohui Zhang
DOI: https://doi.org/10.3969/j.issn.1672-5174.2001.03.011
2001-01-01
Journal of Ocean University of Qingdao
Abstract:The method is based on the reduction of Selenium to volatile SeH2 which is bubbled out by a carrier gas of pure argon and then swept to an Ar-H2 flame quartz atomizer to measure its atomic fluorescence. The hydride generation, transportation, atomisation and some instrumental Parameters were studied by a kind of orthogonal design. The optimum conditions selected are as follows: reactive acidity, 20% HCl; the amount of NaBH4, 0.7% 4.9ml in 5ml sample; the gas flow of Argon, 600m/min; atomize temperature, 200; lamp current, 100 mA; negative voltage, 300V. The detection limit is 0.072μg/dm3. The calibration curve shows an ideal linearity in the concentration range of 0.000~0.1000μg/dm3 Se, The recovery is 95.8%~102.4%.
What problem does this paper attempt to address?