Respiration-Induced Artifacts and Navigator Echo Correction in Gradient Echo Imaging on Human Brain

Fang DONG,Meng-chao PEI,Qian-feng WANG,Hong-wei JIANG,Jian-qi? LI
DOI: https://doi.org/10.3969/j.issn.1000-4556.2014.03.003
2014-01-01
Abstract:Images acquired with gradient echo sequences are susceptible to artifacts induced by frequency offsets varying in spatial and temporal dimensions. This study addressed how respiration-induced local field fluctuations affect the quality of brain images acquired with a three-dimensional multi-echo gradient echo sequence. It was shown that respiration-related artifacts can be corrected by acquire a one-dimensional navigator signal before the application of phase-encoding gradient in each phase-encoding step, which later can be used to correct respiration-related phase shifts during reconstruction to yield artifact-free images. Experimental results showed that both the respiration-related phase oscillations and image artifacts increased with echo time. It was also demonstrated that the phase fluctuations and image artifact levels in both the gradient echo images and T2*maps calculated from mutli-echo data reduced significantly after navigator echo phase correction.
What problem does this paper attempt to address?