Respiratory Self‐gated Multiple Gradient Recalled Echo Sequence for Free‐breathing Abdominal R2* Mapping

Ning Jin,Zhuoli Zhang,Longjiang Zhang,Guangming Lu,Andrew C. Larson
DOI: https://doi.org/10.1002/mrm.22823
2011-01-01
Magnetic Resonance in Medicine
Abstract:Abdominal effective transverse relaxation rate (R2*) mapping is critical for a wide range of applications. However, respiratory motion can lead to significant image quality deterioration and R2* overestimation. For this work, we explored the feasibility of combining respiratory self‐gating techniques with a multiple gradient‐recalled echo sequence for free‐breathing abdominal R2* measurements. In a series of eight normal volunteers, respiratory self‐gated–multiple gradient‐recalled echo methods effectively avoided motion artifacts to produce quantitative R2* measurements in liver, spleen, and kidneys that were comparable to R2* measurements produced while breath‐holding. Respiratory self‐gated–multiple gradient‐recalled echo methods demonstrated the potential to avoid the need for breath‐holding during abdominal R2* mapping. For clinical application, respiratory self‐gated–multiple gradient‐recalled echo approaches could be particularly useful for R2* measurements in those patients unable or unwilling to sustain sufficiently long breath‐holds to avoid motion artifacts. Magn Reson Med, 2011. © 2011 Wiley‐Liss, Inc.
What problem does this paper attempt to address?