Coronary MR Angiography Using Respiration Navigator Echo Trigger

Cheng Liuquan,Gao Yuangui,Ma Lin,Cai Youquan
DOI: https://doi.org/10.3969/j.issn.1005-5185.2006.02.001
2006-01-01
Abstract:Purpose: To investigate the efficiency of respiratory navigator echo gated coronary MRA and the effects of heart rate and rhythm on its imaging quality. Materials and Methods: On 17 healthy volunteers, coronary arteries were imaged using navigator echo gated 3D-SSFP sequence. The diaphragmatic acquisition window was optimized within 3mm using Phase ordering with Automatic Windows Selection (PAWS) technique. The length of the visualized coronary arteries was measured. The correlation of image quality to the heart rate, heart rhythm and respiratory pattern was analyzed. Results: The visualized length of the coronary arteries was RCA 11.6±2.0cm, LM 2.4±0.4cm, LAD 9.0±1.2cm, LCX 5.0±0.8cm. The efficiency of PAWS was 37.7%, taking 6.6±1.0min for one acquisition. The image quality showed no correlation to the respiratory pattern but correlated to heart rate and rhythm, impairing the image quality as they becoming fast and irregular. Conclusion: The PAWS navigator echo gated 3D-SSFP sequence could compensate the respiratory motion for coronary MRA, but it is essential to improve its efficiency to control the motion from the heart.