Radiance calibration of triple-band optical imaging system

Qing HAN,Bang-hui GUO,Jian WANG,Can LI,Jian-zhong ZHANG,Qiang SUN
DOI: https://doi.org/10.5768/JAO201435.0303005
2014-01-01
Journal of Applied Optics
Abstract:T his paper focuses on radiometric calibration of triple-band optical imaging system which includes near ultraviolet (300 nm~380 nm) ,visible (380 nm~760 nm) and near infra-red (760 nm~1 100 nm) ranges .The radiometric calibration system was set up ,the mathe-matical model was established and the radiometric calibration experiment was conducted .As the optical system operated at three wavebands ,technique of sub-band radiometric calibration was adopted .The calibration data were fitted by linear least squares and then the algorithm was modified .At last the calibration data were processed .Experimental results verify the cali-bration accuracy of two methods .The calibration curve obtained by the modified algorithm is with higher accuracy .The relative error of measurements can be less than 5% ,which meets the needs of actual projects .
What problem does this paper attempt to address?