Study of Radiometric Calibration Methods on FTIR Spectrometer

FENG Ming-chun,XU Liang,GAO Min-guang,JIAO Yang,LI Xiang-xian,JIN Ling,CHENG Si-yang,TONG Jing-jing,WEI Xiu-li,LI Sheng
DOI: https://doi.org/10.3969/j.issn.1001-8891.2012.06.012
2012-01-01
Abstract:Infrared radiometric calibration is of critical importance for information quantification of remote sensing of environment at infrared spectrum.In the quantitative analysis,the calibration of the measured spectra is very important.In the paper,there are three specific calibration arithmetic:two points calibration,linear multi-point calibration and nonlinear multi-point calibration.Using Fourier transform infrared spectrometer,the blackbody radiation spectra were measured at different tempetatures.Some of them were used for calibration and one was used to test the result of calibration.The result indicated that linear multi-point point calibration was more accurate.
What problem does this paper attempt to address?