Study on parameter measurement precision of high energy laser beam with large aperture

Qi ZHAO,Qing MENG,Zewei JIANG,Shaoyun HU,Xu GENG,Mingwei GAO
DOI: https://doi.org/10.7510/jgjs.issn.1001-3806.2015.01.020
2015-01-01
Abstract:In order to study the field characteristics of high energy laser, a device for measuring parameters of large-caliber high energy pulsed laser was developed.The device includes a set of autocalibration system that can conduct on-site calibration of its main coefficients.The measuring principle, performance and structure of the device were described, and the measurement error was analyzed by theoretical analysis and experimental verification.The equipment can simultaneously measure the laser energy and intensity distribution.Finally, measurement errors were verified and analyzed by the method of calibration experiments.Measurement errors of laser energy and intensity distribution are 4.5% and 5% respectively. The results show that the equipment has excellent precision, reliability and stability, which could be used in field measurement.
What problem does this paper attempt to address?