Experimental research of spray component concentrations by multi-rainbow spectral-situ measurement with self-calibration method

Kailin CAO,Haoyu JIANG,Can LI,Xuecheng WU,Linghong CHEN,Kunzan QIU,Kefa CEN
DOI: https://doi.org/10.13741/j.cnki.11-1879/o4.2017.05.004
2017-01-01
Optical Technique
Abstract:Calibration of the traditional rainbow measuring technique has the following disadvantages ,complex equip-ment ,precision reliance on manual operation ,poor reproducibility and difficult to adapt to the measurement needs of con-fined space .A self-calibration method based on multi rainbow spectrum is proposed ,eliminating any original rainbow measuring technology calibrate system device .The calibration process is put into the process of rainbow signal process-ing .With the measurement ,it only needs to be processed with spray signal itself .And deviation of water measurement is just 0 .13% .The technology is used in global rainbow technology to measure mix component concentrations in the spray droplets .Factors affecting the issues are explored and the program of two liquid components which involved is optimized . 60% ethanol-water mixture spray is used in the experiments ,and the concentration of mixed multi-component spray is measured in situ measurements .Comparing with the traditional method ,the difference in the measurement results can be controlled within 0 .5% .Moreover ,self-calibration method has better maneuverability in situ measurements .Application of this method shows the potential in related fields .