A New Hand-held Pesticide Residue Monitor Instrument with Temperature Correction Function

Hai Yang,Wei Jin,Linling Wang,Mingwei Su,Yanbo Zheng,Xiaohua Lu
DOI: https://doi.org/10.3321/j.issn:0253-3820.2005.07.036
IF: 1.193
2005-01-01
Chinese Journal of Analytical Chemistry
Abstract:A new hand-held photometric analyzer for pesticide residue monitor based on the acetylcholinesterase (AchE) enzyme catalysis kinetic analysis method, light emitting diodes (LED) as monochromatic source and dry battery as power supply is presented in this paper. This instrument is miniature and intelligent controlled by a single chip microcomputer. Multi-points sampling and light intensity test method are used to improve the measurement precision. In order to eliminate the use of thermostatic apparatus and simplify operation, temperature correction function is first used in this instrument. The performance of this instrument, used improved measuring method was evaluated and compared to national standard measurements with a commercial spectrophotometer, and better relative standard deviation was observed.
What problem does this paper attempt to address?