Study of portable device for pesticide residue detecting with sampling rate emendation

Wang Chao,Zhu Huizhong,Wu Jian,Liang Wudi
DOI: https://doi.org/10.3969/j.issn.1002-7300.2007.10.045
2007-01-01
Abstract:The paper presents an integrated portable device for pesticide residue detection based on the professional standard <The fast determination of organophosphorus and carbamate pesticide residue of vegetable>. In this device transmissive photoelectric sensing module, sampling rate analyzing module as well as thermostatic module is controlled by single chip ADuC812. The photoelectric sensing module can accurately detect the absorbance variation of sampling solution with small module size and low power consuming. To eliminate the operational difficulty of sampling and enhance the work efficiency, sampling rate analyzing parameter calculated on the basis of weight sensor, intelligently get corrected by integrating with absorbance detection value. Thermostatic module supply a 37 ℃ circumstance for the standard detection.