Simultaneous determination of 13 anions in water by bradient elution ion chromatography with suppressed conductivity detection

SONG Wei-de,CHENG Lei,SU Zheng,HUI Xi-dong,CHEN Tai-fa,YANG Li-ming
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2016.0268
2016-01-01
Abstract:A method of gradient elution ion chromatography with suppressed conductivity detection for simultaneous determination of 13 anions in water was established.The effects of flow rate,column temperature,pH value of solution,selection of columns and eluent concentration were studied.Then the suitable gradient elution conditions were found for determination of 13 anions in water.The testing results of 13 anions in water could be more accurate in the condition with the flow rate of 1.00 mL/min,column temperature of 35℃ and the value of solution pH in the range of 5.6 ~6.9.ExcePt the unstable carbonate ion,the linear relation of other 12 anions were good in the range of 0.005 ~ 20.0 mg/L and the linear correlation coefficients were not less than 0.9995 with the limits of detection (S/N =3) of 12 anions in water of 0.8 ~ 16.0 μg/L.The average recoveries of the twelve anions in water were 90.4% ~ 107.6% in the levels of 0.10,0.20 and 5.00 mg/L,and the relative standard deviations (RSDs)were in the range of 0.08% ~ 5.3%.This method has the virtues of simplicity,quickness,good sensitivity and high accuracy,and is absolutely suitable for trace analysis and determination of 13 anions in drinking water,ground-water,irrigate-water and other environmental water.
What problem does this paper attempt to address?