Determination of disinfection by-products in water by ion chromatography

W. Jian
Abstract:Objective:To establish a new conductivity detection ion chromatography determination method for contents of BrO-3,ClO-2,ClO-3,DCAA,TCAA in water by the large volume direct injection.Methods:In the suppressed conductivity detection,Dionex IonPac AS19 column was used as separation column,sample injection volume at 500 μl,with KOH gradient elution generated by on-line eluent generator.Column temperature was 25℃~26℃.DBPs and conventional anion in water were detected in 35 min at the same time.Results:The DBPs content showed good linearity within the concentration range of 5 μg/L~2000 μg/L.Limit of determination was low.Relative standard deviations(RSD) was less than 3%,recoveries were between 84%~117%.Conclusion:The method is simple,practical and reliable.
Environmental Science,Chemistry