Determination of Trace Methanol in Perfume Samples by Gas Chromatography-Flame Ionization Detector

Yun LIU,Zhi-quan LI,Man HE
DOI: https://doi.org/10.13526/j.issn.1006-6144.2018.01.023
2018-01-01
Abstract:A method based on gas chromatography-flame ionization detector(GC-FID) was proposed for the quantification of trace methanol in perfume samples.The sample was directly injected into GC for subsequent analysis,and the false positive phenomenon was avoided by choosing appropriate separation column and optimizing heating program.The absolute detection limit of the proposed method was 5 ng for methanol,and a good linear relationship was obtained in the concentration range of 50-500 mg/mL.The proposed method was applied for the analysis of trace methanol in five perfume samples,and the recoveries of 98.8%-103% were obtained for three spiking level(400 μg,2 750 μg,4 000 μg).The method avoids the false positive results which commonly occurs in conventional methods to a great extent,and it is sensitive,accurate and fast,with a good application potential in the analysis of methanol in aromatic cosmetics(perfume).
What problem does this paper attempt to address?