Calibration Method of Fluorescence Spectrometer and Evaluation of Uncertainty

Yan-jie HUANG,Cheng-hui LI,Jin-yan ZHOU,Zhen-yu HUANG,Ling CHEN,Xin WANG,Yan GUAN
DOI: https://doi.org/10.13822/j.cnki.hxsj.2019006931
2019-01-01
Abstract:Fluorescence spectrometer is a common analytical instrument, which is suitable for the analysis in food, health, agriculture and environment areas.It can provide one or more information such as excitation spectrum, emission spectrum, fluorescence intensity, quantum yield, fluorescence lifetime, and fluorescence polarization. The calibration methods and uncertainty evaluation methods for fluorescence spectrometer with detection wavelength ranging from 200 nm to 1 700 nm are introduced, which containing wavelength indication error and repeatability, detection limit, measurement linearity and peak intensity repeatability.The expanded uncertainty of wavelength indication error is 0.20 nm (k = 2); the limit of detection is 8. 09×10-11 g/m L with an expanded uncertainty to be 3.60×10-11 g/m L (k = 2); the relative expanded uncertainty of measurement linearity is 3.1% (k = 2); the relative expanded uncertainty of peak intensity repeatability is 2.0% (k = 2).The calibration process and result evaluation of fluorescence spectrometer described will provide a good example for calibration of related instruments.