The regular daily quality control method for the new semiconductor cardiac SPECT

张宗耀,郭风,任俊灵,吴大勇,汪蕾,何作祥,许百灵,方纬
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2016.11.008
2016-01-01
Abstract:Objective:To establish a regular daily quality control method for the new cardiac semiconductor SPECT. Methods: The 740 MBq 57Co flood source for NM Discovery 530c was used. Quality control parameters and reference value were energy peak 122±1.5 keV, FWHM≤7.5%, uniformity≤9%, central bad pixels≤30, the number of detectors with more than 7 bad pixels ≤3. The 370 MBq 57Co line source for Spectrum D-SPECT was used. Reference value of quality control parameters were regional homogeneity index >90%, global homogeneity index>85%, effective field of view>90%, specificity 2277.6±38.1cpm/mCi. Results:10 consecutive daily quality controls were performed for NM Discovery 530c and Spectrum D-SPECT. 10 cases of 99Tcm-MIBI myocardial perfusion imaging per day were acquired separately and the quality of images was evaluated by experienced nuclear medicine physicians. All quality control parameters for NM Discovery 530c and Spectrum D-SPECT were acceptable and the excellence of images was 100%. Conclusion:The standard daily quality control method for the new semiconductor cardiac SPECT can guarantee the image quality of myocardial perfusion imaging.