Clinical contrast studies of domestic and imported color doppler ultrasound in prenatal ultrasonography

张利,郭丽娟,崔立刚,张明博,王红燕,杨敬英,马相国,郭亚美,宫俊英
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2016.12.013
2016-01-01
Abstract:Objective:To discuss whether application of domestic color Doppler ultrasound in prenatal ultrasonography in different level hospitals is feasible by clinical contrast studies of domestic and imported color Doppler ultrasound in seven different level hospitals.Methods: To make comparative evaluation of images of fetus in prenatal ultrasonography in the seven hospitals and give a mark.Results: Each hospital has 30 cases, so there are 210 cases in 7 hospitals. The evaluation results of four domestic color Doppler ultrasound instruments are 87.75 (No. 1 instrument), 96.20 (No. 2 instrument), 81.31 (No.3 instrument) and 84.90 (No.4 instrument). The result of imported instrument is 105.0. Compared with the imported instrument, the result of four domestic color Doppler ultrasound is p<0.05. So there is statistical significance (t=3.03,t=3.18,t=2.90,t=3.23;P<0.05).Conclusion: Domestic color Doppler ultrasound can meet the needs of prenatal ultrasonography and other regular work in primary hospitals and second-class hospitals, but it cannot meet the needs of prenatal ultrasonography in third-class hospitals. So it needs to be improved.