A comparison study of harmonic imaging and fundamental imaging in fetal echocardiography

赵博文,寿金朵,徐海珊,吕江红,范妙英,汤富刚,范晓明
DOI: https://doi.org/10.3760/j.issn:1004-4477.2002.11.009
2002-01-01
Abstract:Objective To directly compare the quality of harmonic imaging (HI) and fundamental imaging (FI) in fetal echocardiography and to determine any differences in image quality between the two modalities. Methods Fetal echocardiograms were performed by using FI and HI in 58 fetuses, image quality and visualization of left and right atria, left and right ventricles, mitral and tricuspid valves, aortic and pulmonary valves, left and right ventricular outlets were evaluated and compared between FI and HI.Results Mean HI scores were higher than mean FI scores for all the cardiovascular structures ( 2.73 ± 0.43 vs 2.16 ± 0.69 , P 0.001 ). Compared with FI, HI improved the image quality and visualization of fetal cardiac structures in subjects with good ( 2.73 ± 0.43 vs 2.88 ± 0.32 ,P 0.001 ) and suboptimal ( 1.65 ± 0.41 vs 2.58 ± 0.47 ,P 0.001 ) echocardiographic windows. The interobserver correlation coefficient for the grading scores was 0.74  (P0.001 ). Conclusions Harmonic imaging enhances and improves the image quality of fetal echocardiography, it has important potential role in cardiac imaging in the fetus.
What problem does this paper attempt to address?