Results and analysis of participation in the medical accelerator dose output comparison organized by IAEA/WHO

Hongfu ZHAO,Zhipeng ZHAO,Zhiqiang WANG,Guanghui CHENG
DOI: https://doi.org/10.13491/j.issn.1004-714x.2019.01.019
2019-01-01
Abstract:Objective To check the accuracy of medical accelerator dose by participating in international comparisons by postal TLD organized by the IAEA/WHO.Methods Apply for the IAEA/WHO postal TLD dose audit to obtain the participation qualification. For each energy condition, irradiate the 2 TLD samples at an irradiation dose of 2 Gy and fill out the data sheet. Mail the TLDs and data sheet back to the IAEA lab, and the results were given after the IAEA measurements.Results The mean measured dose of the 6 MV photon beam was 2.00 Gy with an average deviation of 0.0%, and the doses for 2 TLDs irradiated at 10 MV photon beam were both 2.02 Gy with an average deviation of-0.8%. Conclusion The comparison results met the requirement within allowable deviation of ±5% set by IAEA.