Xiaoshen WANG,Xiangkun DAI,Gaoxiang CHEN,Xiaohu CONG,Jinyuan WANG,Mingyong SONG,Shouping XU
DOI: https://doi.org/10.3969/j.issn.1005-202X.2017.04.004
2017-01-01
Abstract:Objective To investigate the dose-response non-uniformity of self-developing EBT3 film and its influence on the performance of the film for dose quality assurance of tomotherapy plan.Methods Ten EBT3 films from the same batch were selected for analysis.Each film was divided into 8 areas,and after exposure to irradiation of different doses,the doseresponse non-uniformity of the film was assessed,and the measured and corrected dose-response curves A and B were created.Ten intensity-modulated radiotherapy (IMRT) plans from the TomoHD system were selected for dose verification using EBT3 films,and the gamma passing rates were analyzed with the dose-response curves A and B.Results The nonuniformity of the pixel values in the 10 films was analyzed,and the coefficients of variationof the pixel values in each film were 2.32%,1.32%,0.74%,1.20%,0.87%,0.72%,0.88%,0.72%,1.00%,and 0.70%.Analysis of the 10 IMRT plans with the dose-response curves A and B showed that the passing rates of the films differed significantly between the two groups for gamma values of 1.5 (P=0.007) and 1.0 (P=0.008) but not for a gamma value below 1.0.Conclusion The selfdeveloping films showed differences in the dose-response non-uniformity,which was also related inversely with the exposure dose of the films.Correcting the EBT3 dose-response curves is necessary in clinical dose verification of the tomotherapy plans,especially for the low dose areas.