Energy effect on the gamma analysis in the proton therapy end-to-end test with EBT film

Seung Hoon Yoo,WeiSing Tan,Yen Hwa Lin,Joerg Rottman,Jungmo Yoo,EunHo Kim
DOI: https://doi.org/10.1007/s40042-023-01001-7
2024-02-04
Journal of the Korean Physical Society
Abstract:We analyzed the effect of energy on the gamma analysis results in an end-to-end test using EBT3 film for the proton therapy facility commissioning. The EBT3 film calibration was performed using a single energy layer and SOBP-based film calibration methods to compare which is better. The results showed that the SOBP method exhibits better results for gamma passing ratio with all criteria. Also, results in gamma values showed the best score for middle energy level for both methods. This is due to the increased film sensitivity with averaged LET from SOBP and maximum dose with middle energy proton beams. All gamma evaluation results showed over 95% with 3%-3 mm criteria with SOBP method; however, in lower single energy-based calibration method, gamma result can be worse than 95%, but still over 90% of passing criteria. Dose measurement results showed the uncertainty less than 3% compared to calculated dose from the treatment planning system. SOBP-based calibration can be suggested for more accurate end-to-end test for the commissioning.
physics, multidisciplinary