Incidence and related factors of administration events in intermediate frequency electrical therapy

杨飞,刘斌,孙凤,陶庆梅,徐荣彬,杨俊,翟伟,钟蕾,刘东红
DOI: https://doi.org/10.3969/j.issn.1001-1242.2017.09.007
2017-01-01
Abstract:Objective:To investigate the incidence and related factors of medical device administration events (MDAE)with intermediate frequency electrical stimulator's.Method:The medical device administration events' prevalence survey of 63 intermediate frequency electrical stimulation in 8 hospitals in Beijing was conducted in two month,.Device state,maintenance situation,patients' physical characteristics,clinic operation situation and running state were recorded,then the logistic regression models were used for single factor and multifactor analysis.Result:In the collected 3075 records,55 cases of MDAE happened,the percentage was 1.8%.Replacing the electrodes based on damage or unusable was more dangerous than those based on using time,OR=7.383,95% CI:l.661-32.819,P=0.009;.Non-using moist mat under the electrodes was 3.966 times more dangerous than those using ones,95%CI:1.707-9.215,P=0.001.Using the electrodes without liquid medication was safer,OR=0.307,95%CI:0.148-0.636,P=0.001.Conclusion:Related factors of intermediate frequency electrical stimulator's administration events (MDAE) were found,and risk control suggestions were proposed in this work.These will improve the clinic use security,and guarantee the patients' life safety in intermediate frequency electrical therapy.