Abstract:Objective:To predict the interface pressure between residual limb and prosthetic socket for below-knee (BK) amputees during walking.Method:A quasi-dynamic finite element model was built based on the actual geometry of residual limb, internal bones and socket liner. To simulate the friction/slip boundary conditions between the skin and liner, the automated surface-to-surface contact was used. Not only the variable external loads and material inertia but also the coupling between the large rigid displacement of knee joint and small elastic deformation of residual limb were considered.Result: The interface pressure was predicted in one gait cycle. The high pressures fell over popliteal depression, middle patella tendon, lateral tibia and medial tibia regions. The interface pressures except for the one at patella tendon region had the similar double-peaked waveform shape of ground reaction force in stance phase. Conclusion:The quasi-dynamic finite element model built in this paper can almost really simulate the motion of knee joint during walking, it can be used for prediction of interface biomechanical property on residual limb in a prosthetic socket and digital design of modern prosthesis.