Feasibility and parameter optimization of conventional magnetic resonance imaging for nude ;mice

Jia CHEN,Ming-fu GONG,Hua YANG,Wei-bin WANG,Yun LIU,Liu YANG,Tong-sheng SHU
DOI: https://doi.org/10.3969/j.issn.1672-6170.2016.03.037
2016-01-01
Abstract:Objective To explore the feasibility of magnetic resonance imaging(MRI)for nude mice using conventional MR scanners.Methods Liver MR images of nude mice were obtained using a small joint coil and a dedicated small animal coil.The image quality of the coils was evaluated according to the artifact,uniformity and the contrast of images.T1 mapping and T2 mapping were per-formed to nude mice livers to determine T1 value and T2 value.Then,the echo time(TE)and repetition time(TR)were determined ac-cording to the T1 value and T2 value,Finally,the scan sequence was optimized by adjusting other parameters.Results In FSE T1 WI, SE T1 WI,FSE T2 WI,GRE T2*WI sequence,the image scores of small joint coil and dedicated small animal coil were 2.6 ±0.54,3.2 ±0.84,2.6 ±0.89,2.4 ±0.55 and 3.6 ±0.55,3.8 ±0.45,3.2 ±0.84,2.6 ±0.55,respectively.SNR of small animal coil was significantly higher than that of small joint coil(P<0.05).After optimization,the image resolution,contrast and uniformity significantly were improved.Conclusion MRI in vivo nude mice could be performed on conventional MR scanners.It provides a reference for the general research institutions in small animal MR imaging.