Zero-fluoroscopic implantation of pacemaker guided by Ensite-NavXTM

王炎,陈光志,姚焰,郭平,薛玉梅,吴书林,汪道文
DOI: https://doi.org/10.3969/j.issn.1001-9057.2016.06.006
2016-01-01
Abstract:Objective To explore the feasibility and safety of non-fluoroscopic implantation of pacemaker guided by Ensite-NavXTM.Methods Patients with severe bradycardia were arranged for zero-fluoroscopic implantation of pacemaker guided by Ensite-NavXTM when C-arm X-ray machine was being used for other procedures.We reconstructed the model of right atrium,right atrial appendage,and right ventricle with Ensite-NavXTM.Subclavian vein was punctured.Atrial or ventricular leads were collected to three dimensional mapping system and cables for examining pacing parameters.The leads were placed into the target chambers.Ventricular leads were further advanced 3-5 cm after reaching the ideal position. Atrial leads or ventricular leads were fixed when the pacing parameters were still satisfactory after gently inserting and retracting the leads.Results Five cases underwent implantation procedures guided by three-dimensional electro-anatomic mapping system and without X-ray.Three cases were implanted with single-chamber pacemaker and two cases with dual-chamber pacemaker.The total fluoroscopic time was 0 minute.The average procedure time was (51 .3 ±1 3.1 )minutes and (97.5 ±1 6.3 )minutes for single-chamber and dual-chamber pacemakers,respectively.The pacing parameters were normal during follow up.Conclusion Single-chamber or dual-chamber pacemakers can be safely and effectively implanted under the guidance of three-dimensional electro-anatomic mapping system.Our method offered a choice for some special population who need to avoid radiation or in the occasion when X-ray machine is not useable.
What problem does this paper attempt to address?