Comparative study on measuring method for initial extension force of elastomeric auxiliaries for use in orthodontics

Yuan LI,Hong LIN,Gang ZHENG,Yong-xiang XU
DOI: https://doi.org/10.3760/cma.j.issn.1002-0098.2013.s1.026
2013-01-01
Abstract:Objective To study on clinical applicability of the two methods by comparing the initial extension force measured by relax extension & delay force measurement and direct force measurement.Methods Four kinds of orthodontic elastics,three kinds of ligatures,three kinds of elastomeric chains and three kinds of dumbbell force modules were used to measures the initial extension force by the two test methods.All these elastomeric auxiliaries had different size and manufacturer.Using independent sample t-test to compare the differences of initial extension force measured by two methods.Calculate the contrast ratio with the formula:initial extension force measured by direct force measurement divided by initial extension force measured by relax extension & delay force measurement.Results The data of the initial extension force measured by direct force measurement was more than by relax extension & delay force measurement and the two methods had significant differences (P < 0.05) and the contrast ratio was 1.1 times to 4.8 times.The contrast ratio of orthodontic elastics which were made of natural rubber was 1.1 times to 1.9 times,and was less than ligatures (1.9 times to 3.2 times),elastomeric chains (2.4 times to 4.5 times) and dumbbell force modules (2.8 times to 4.8 times) which were made of polyurethane rubber.Conclusions The data of the initial extension force measured by the two measurements had significant differences.Using relax extension & delay force measurement can reproduce the operation process using elastomeric auxiliaries by dentists in orthodontic clinic.
What problem does this paper attempt to address?