Evaluation of Protrusive Occlusal Discrepancies Resulting from Measuring Accuracy by Mathematical Methods

LI Hong-bo,FENG Hai-lan
DOI: https://doi.org/10.3969/j.issn.1009-3761.2006.01.002
2006-01-01
Abstract:Objective: To evaluate the occlusal discrepancies resulting from measuring errors by the ARCUS digma system in acquiring articulator parameters. Methods: First we deduced the mathematical relationships between protrusive occlusal discrepancies and different condylar inclinations or incisal guide angles on articulator. And then the protrusive occlusal discrepancies induced by the maximum measured value, the minimum measured value and the mean for a preset value by the ARCUS digma system were calculated. Results: When the condylar inclinations were smaller than 30 degrees, the protrusive occlusal discrepancies induced by both the maximum and minimum measured values were within 0.2mm. Whereas the discrepancies induced by the maximum and minimum measured incisal guide angles were a little larger. If the condylar inclinations got to 45 degrees or the incisal guide angles got to 80 degrees, the occlusal discrepancies were notable. Conclusions: By the ARCUS digma system, we could acquire articulator parameters which led to neglectable protrusive occlusal errors. We could adopt those condylar inclinations which were smaller than 30 degrees just by measuring for one time, while the incisal guide angles should be measured for several times to get the means. If the condylar inclinations exceed 30 degrees or the incisal guide angles exceed 70 degrees, we should be cautious in results adoption.