Low dose in contrast-enhanced CT scan of liver:80 kVp with sinogram affirmed iterative reconstruction technique

姜健,王可,许玉峰,徐莎莎,刘建新,王霄英
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2016.04.008
2016-01-01
Abstract:Objective:To evaluate the feasibility of 80kVp with sinogram affirmed iterative reconstruction (SAFIRE) technique in liver contrast-enhanced CT scan of normal weight adults.Methods:Forty-six patients with body mass index of 18~24kg/m2 and suspected of hepatic diseases were prospectively recruited and then underwent liver contrast-enhanced CT using low dose (80kVp and 420mAs)and personalized contrast injection protocol.Filtered back projection (FBP)and five SAFIRE algorithms were used respectively for image reconstruction,and 6 sets of images were obtained for each patient. For objective evaluation,image noise,contrast-to-noise ratio (CNR)and figure of merit (FOM)were measured and calcula-ted.For subjective evaluation,subjective score of image noise,vessel conspicuity,artifacts and overall image quality were as-sessed,as well as the timing of late arterial phase.Results:For objective analysis,images with SAFIRE-5 algorithm showed the least image noise and the highest CNR and FOM,with statistical difference from those of the other 5 groups (P<0.05).For subjective analysis,images with SAFIRE-5 algorithm showed the highest scores of image noise,vessel conspi-cuity and beam hardening artifacts,and the least score of reconstruction artifacts.As for overall image quality,the image scores of SAFIRE-3 were the highest compared with the other 5 groups.95.6% (44/46)of the patients could obtain appro-priate images in late arterial phase.Conclusion:In liver contrast-enhanced CT,80kVp with iterative reconstruction technique and personalized contrast media inj ection protocol can obtain high quality post-contrast images;SAFIRE-3 algorithm is recommended for image reconstruction.
What problem does this paper attempt to address?