Evaluation Of Dose Reduction And Image Quality In Chest Ct Using Adaptive Statistical Iterative Reconstruction With The Same Group Of Patients

L-P Qi,Y. Li,L. Tang,Y-L Li,X-T Li,Y. Cui,Y-S Sun,X-P Zhang
DOI: https://doi.org/10.1259/bjr/66327067
2012-01-01
British Journal of Radiology
Abstract:Objectives: The objective of this study was to compare the image quality and radiation dose of chest CT images reconstructed with a blend of adaptive statistical iterative reconstruction (ASIR) and filtered back-projection (FBP) with images generated using conventional FBP.Methods: Patients with chest CT re-examinations were alternately assigned to two scanners with different reconstruction techniques. The study groups included noise index (NI) 11 with 30% ASIR (A30), NI 13 with 40% ASIR (A40), NI 15 with 50% ASIR (A50) and NI 17 with 60% ASIR (A60), sequentially changed every 2 months. The control images were obtained using FBP and NI 11. All acquisitions were performed with automatic dose modulation. Paired t-test and non-parameter test were applied to compare the difference.Results: The radiation doses were significantly lower in the examinations that used ASIR (p < 0.001). The mean dose reduction rate was 27.7%, 45.2%, 57.1% and 71.8% for Groups A30, A40, A50 and A60, respectively. The image quality of Groups A30-A50 was not inferior to that of the control examinations. The image noise of Group A60 was greater and subjective image quality was inferior to that of the control.Conclusions: ASIR enabled the use of a higher NI with automatic dose modulation. With 50% ASIR and a NI of 15, the effective radiation dose was reduced by 57%, without compromising image quality.