Evaluation of iterative AIDR and filtered back projection image algorithm on coronary computed tomography angiography

严诚,陈刚,丁玉芹,杨雪,陆秀良,张利军,曾蒙苏
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2014.06.021
2014-01-01
Abstract:Objective:To compare the image qualities of coronary computed tomography angiography (CCTA)using adaptive iterative dose reduction (AIDR)and filtered back projection (FBP)algorithm and to evaluate the clinical value of AIDR.Methods:Twenty-six patients with suspected stable angina underwent regular radiation dose scanning protocol on 320 MDCT scanner.The collected raw data were then reconstructed with AIDR and FBP respectively.CT value and back-ground SD value were measured at the aorta root,proximal and middle right coronary artery,left main coronary artery,prox-imal left anterior artery,as well as proximal left circumflex artery.SNR and CNR were also calculated.Two observers com-pared the quality of the images with subjective grading.Results:Twenty-six patients completed CCTA within one heart beat.The average CT attenuation measured at segments of coronary arteries was not statistically different (P>0.05).Back-ground SD value,SNR and CNR calculated were statistically different (P<0.001).The subjective grading of image qualities between two groups were 3.73±0.49 for AIDR and 3.59±0.61for FBP,respectively,and was statistically different (P<0.01).Conclusion:The quality of images reconstructed with AIDR was better than that of FBP.AIDR can reduce the ima-ging noise significantly,improve the image quality and help reducing radiation dose potentially.