Feasibility of contrast-enhanced spectral mammography in the diagnosis of breast cancer

张承中,王庆国,王建丰,何之彦,张贵祥,李康安
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2014.12.018
2014-01-01
Abstract:Objectire:To investigate the feasibility of contrast enhanced spectral mammography (CESM)in the diag-nosis of breast cancer by using a comparative study with MG and CE-MRI.Methods:After ethical approval,63 women with suspected breast cancer underwent CESM,MG and CE-MRI.MG and CESM were performed before and after injection of nonionic iodinated contrast agent.CE-MRI was performed using a VIBRANT sequence after injection of Gd-DTPA.All the images were analyzed by three experienced breast radiologists,and the maximum dimension of lesions were measured.The diagnostic results of three imaging methods were compared to pathological results.Results:Of 63 cases,there were 42 malig-nant lesions and 21 benign lesions.The diagnostic accuracy of CESM、CE-MRI and MG for breast lesions was 66.67% (42/63),65.08% (41/63)and 57.14 (36/63),with no significant difference among the three methods (P >0.05);in diagnosis of malignant breast lesions,the sensitivity of CESM was high (100%),but the specificity (42.9%)was low.The average maximum dimension of the breast cancers measured by CESM,MG and CE-MRI was (28.33±17.24),(26.50±16.21)and (26.45±15.94)mm,with no significant difference (P >0.05)when compared with the pathological results respectively. Conclusion:Our Preliminary results show that the diagnostic accuracy of CESM in detection of breast cancer is excellent and higher than that of MG,so it may be an important diagnostic method for breast lesions.