The clinical application of Amplatzer vascular plug Ⅱ in treating perimembraneous ventricular septal defects

Ying GUO,Ting-liang LIU,Zhi-qing YU,Mei-rong HUANG,Wei GAO
DOI: https://doi.org/10.3969/j.issn.1008-794X.2014.04.003
2014-01-01
Abstract:Objective To discuss the feasibility of using Amplatzer vascular plug Ⅱ to occlude perimembraneous ventricular septal defects (VSD). Methods Transcatheter closure by using Amplatzer vascular plugⅡwas carried out in 5 patients with perimembraneous VSD. The patients included 3 girls and 2 boys, with a mean age of (3.8 ± 1.3) years (ranged 3-6 years) and a body weight of (11.8-22.0) kg. The vascular plug was released after its position was confirmed by angiography and echocardiography. After the treatment, left ventricular and aortic angiography as well as color echocardiography was performed to evaluate the therapeutic results. Each patient underwent electrocardiographic and echocardiographic examination at 24 hours after closure, at the time of discharge from the hospital, and at 1, 3, 6 month during the follow-up period. The results were analyzed. Results The echocardiogram and angiogram showed that the diameter of VSD was 6.4 - 9.0 mm in the left ventricle side with an aneurysmal transformation and was 2.1 - 2.9 mm in the right ventricle side. The distance from the upper rim of VSD to aortic valve was 2.1 - 3.8 mm. Transcatheter deployment of the device was successfully accomplished in all patients. The size of deployed device ranged from 4 to 10 mm. Angiogram and echocardiogram revealed that the defect was successfully occluded in all patients except one who showed a trivial residual shunt at the lower rim of the device, and the residual shunt disappeared one month later. During follow-up period, no patient developed bundle branch block or atrioventricular block. The position of the device remained stable. Conclusion Transcatheter closure of perimembranous ventricular septal defects, especially in the patients with long-tube type aneurysmal transformation, the use of Amplatzer vascular plug Ⅱ as the occluder is safe and effective.(J Intervent Radiol, 2014, 23:284-287).
What problem does this paper attempt to address?