Reconstruction and Numerical Calibration of Electron Density Images of Medical Dual-energy CT Based on Δμ-ρe

Jiaojiao Qin,Fengrong Sun,Fanglei Liu,Peipei Jin,Haitao Yu,Hai Zhong,Guihua Yao
DOI: https://doi.org/10.16289/j.cnki.1002-0837.2017.04.007
2017-01-01
Abstract:Objective To obtain precise electron density images,the electron density images of dual energy CT are reconstructed and numerical calibration is made.Methods Based on the principle of basic material decomposition imaging of medical dual-energy CT (DECT),the theoretical electron density image was calculated and the monochromatic image of X ray linear attenuation coefficients under two different energy levels was acquired.Then,the combination of X ray linear attenuation coefficients under the high and low energy levels was linearly fitted with the theoretical electron density values (Δμ-ρe);Meanwhile the corresponding goodness of the fitting was calculated.Finally,the reconstruction and numerical calibration of electron density image of medical DECT were realized by virtue of the best goodness of fitting.Results The simulations and the quantitative analyses indicated that the pixel values of electron density reconstruction image obtained by the proposed method were more precise and the image quality was much higher.The practical data experiments showed the effectiveness of this method.Conclusion This paper presents a practical and accurate method for reconstruction and numerical calibration of electron density images of medical DECT.