The Correlative Study on Image Quality , Concentration of Iodinated Contrast Material and Radiation Dose

高凯,孙立军,邓振生,牟轩沁,余厚军,张学昕
DOI: https://doi.org/10.3969/j.issn.1002-1671.2005.01.001
2005-01-01
Abstract:Objective To study the correlation among image quality , concentration of iodinated contrast material , and radiation dose.Methods Radiographic images were obtained of a phantom which was 1.2 mm diameter vessels containing iodine at concentration between 4.0 and 50 mg/ml. The detection threshold concentration of iodine was determined by inspecting radiographic images at a range of X-ray tube voltages and input exposure levels. Surface doses were obtained from measured X-ray tube output data. The correlative changes were observed among X-ray tube voltage, surface dose, air kerma , and the detection threshold concentration of iodine. The relationship between the pixel values of images and X-ray tube voltage was analysed with a different situation of air kerma.Results Surface doses were found to be decreased with increasing the X-ray tube voltage when air kerma was varied between 1.1 μGy/frame and 4.0 μGy/frame. The pixel value was kept relatively constant with air kerma of 4.0 μGy/frame and increased with air kerma of 1.1 μGy/frame when X-ray tube voltage was increased. The most vessels were showed with the X-ray tube voltage of 85 kVp.Conclusion Radiation injuries may effectively decreased with reducing X-ray tube voltage and exposure level under the precondition of X-ray image quality to be guaranteed.