Chiral separation of aranidipine enantiomers by HPLC

Shu-bo DONG,Han-yue YANG,Xue-min CHEN,Zheng-yu YAN
DOI: https://doi.org/10.16155/j.0254-1793.2017.06.25
2017-01-01
Abstract:Objective:To establish an HPLC method to separate aranidipine enantiomers for optical purity detection of S-aranidipine.Methods:The method was performed on a Chiralcel OJ-H column [cellulose tri (4-methylbenzoate) chiral stationary phase].The mobile phase was n-hexane-ethanol (75 ∶ 25) at a flow rate of 1.0 mL· min-1.The column temperature was 35 ℃ and the detection wavelength was 238 nm.Results:The resolution between the peaks of S-aranidipine and R-aranidipine was 3.5.The method was found to show good linearity over the concentration range of 0.04-2.0 μg· mL-1.The linear equation was A=61 971C-0.321 4,and the regression coefficient was 0.999 9.The limit of quantification and limit of detection of R-aranidipine were 0.8 ng and 0.3 ng,respectively.The recoveries of R-aranidipine (n=3) were 97.5%,98.5%,and 99.8% respectively at the concentration level of low,medium and high,with the RSD (n=9) of 1.5%.The test solution was stable for at least 12 h at 25 ℃ away from light.The R-aranidipine content in S-aranidipine sample was 0.08%.Conclusion:With good resolution and repeatability,the method can be reliably used for chiral separation of aranidipine,and for determination of the optical purity of S-aranidipine.
What problem does this paper attempt to address?