Comparison of EMIT and 2D-HPLC-UV in the Quantification of Van-comycin in Human Plasma

Min WANG,Hang LIU,Lu JIN,Feng WANG,Wei-hong GE
DOI: https://doi.org/10.13664/j.cnki.pcr.2018.01.008
2018-01-01
Abstract:There exist great individual differences in the concentration of vancomycin, so it is necessary to monitor its concentration in clinic. An enzyme multiplied immunoassay technique (EMIT) and a two-di-mensional liquid chromatography coupled with ultraviolet detection method (2D-HPLC-UV) were established to measure the concentration of vancomycin in human plasma. The linear range, accuracy and precision of the two methods satisfied the requirements for bioanalysis. Sixty plasma samples from patients intravenously infused with vancomycin were detected by EMIT and 2D-HPLC-UV, respectively. There were no significant differences in the plasma concentrations of vancomycin detected by EMIT and 2D-HPLC-UV. A good cor-relation was obtained between the two methods. The regression equation was Y2D-HPLC-UV=0.9670×XEMIT-0.7029. Bland-Altman plots analysis showed the two methods were consistent. Both EMIT and 2D-HPLC-UV were effective and reliable in the quantification of vancomycin. The monitoring purpose, laboratory conditions and staffing should be taken into consideration when a method was chosen for monitoring the concentration of vancomycin.