Simultaneous detection of concentrations of imatinib and voriconazole in human serum by RP-HPLC

Xing-xian LUO,Lin HUANG,Tai-feng LI,Xue-cai XUE,Yu-hang WANG,Yue CHEN,Chang-qing YANG,Wan-yu FENG
DOI: https://doi.org/10.13286/j.cnki.chinhosppharmacyj.2018.01.09
2018-01-01
Abstract:OBJECTIVE To establish a method for determining the concentrations of imatinib and voriconazole in plasma by HPLC.METHODS Blood samples were pretreated by using ethyl acetate-n-hexane (75 ∶ 25) with carbamazepine as the internal standard,then concentrated by nitrogen and detected by HPLC after dissolution.Analytical column was ZORBAX SB-C18 (250mm × 4.6 mm,5 mm),mobile phase was methanol-acetonitrile-ammonium acetate buffer (0.025 mol· L-1 CH3 COONH4,pH value was adjusted to about 4.5 by CH3COOH)=20∶ 32∶48(V/V/V) at a flow rate of 1.0 mL·min-1.Detection wavelength was set at 264 nm and temperature was kept at 35 ℃,and the injection volume was 20μL.RESULTS The calibration graphs were linear in range of 0.10-5.00 μg·mL-1 for imatinib and 0.10-6.00μg·mL-1 for voriconazole (R =0.998 and 0.999,respectively).Limits of quantification were 0.1 μg·mL-1 both for imatinib and voriconazole.Extraction recoveries of low,medium and high concentrations were 69.21%,71.23 %,73.53% for imatinib and 76.23%,78.12%,79.34% for voriconazole,respectively.Intra-and inter-day RSD values were less than 15%.CONCLUSION The method is simple,quick and sensitive for drug monitoring and pharmacokinetics study on imatinib and voriconzole.
What problem does this paper attempt to address?