Comparative study on radiation dose and image quality between spectral imaging & conventional CT plain in thorax

Qiu-zhen XU,Hai-xue ZHU,Yan LV,Hai-lei SHAO,Xiao-hui CHEN,Sheng-hong JU
DOI: https://doi.org/10.3969/j.issn.1671-6264.2017.05.031
2017-01-01
Abstract:Objective:To explore the difference between the radiation dose and image quality of chest spectral CT imaging and conventional chest CT scan .Methods:33 adult patients diagnosed or suspected with primary broncho-genic carcinoma prospectively underwent non-enhanced CT scan and post-contrast CT scan with spectral imaging scanner .Both types of the scan evaluated the noise objectively , then scored them subjectively through adipose tissue as well as musculature on various levels of the chest .The radiation doses were recorded at the same time .The radi-ation doses of both types of scans were compared by independent samples T -test, which was also used to compare the objective measures and subjective scores between the 70 kev-monochromatic energy image and mix-energy rou-tine scan.Results: There was no significant difference in the CT dose index and the effective dose between the spectral scanning and conventional scanning ( P>0 .05 ) .There was also no significant difference of image of noise and image quality for the adipose tissue and musculature among the three groups with different levels of 70 kev-mon-ochromatic energy image and mix-energy routine scan (P>0.05).Conclusion:Firstly, radiation doses of spectral chest CT is correspondent to routine CT .Secondly , the noise of 70 keV monochromatic energy imaging is lower than that of routine scanning FBP reconstructed images , which means it can be routinely conducted in clinic .